Select Your Printer Model
(Default: k-605 DTF Printer)
Choose your model from the drop-down menu below.↓↓↓
Select Your Printer Model
(Default: k-605 DTF Printer)
Choose your model from the drop-down menu below.↓↓↓
Applicable model: LINKO K605 DTF Printer
This guide is used to solve vertical misalignment or ghosting between the white ink layer and the color ink layer.
Calibration is needed when the white ink layer and the color ink layer are not aligned vertically in the printed image.
| Condition | Check Result | Next Step |
|---|---|---|
| White ink layer and color ink layer are fully aligned | Normal | No calibration needed |
| Vertical misalignment or ghosting between white ink layer and color ink layer | Abnormal | Do step calibration first. Then do vertical calibration |
| Broken nozzle lines, missing nozzles, obvious missing colors, or material slipping | Direct calibration is not recommended | Check nozzle status, material, and film feeding first |
Important: Follow this sequence:
Pre-check → Step calibration → Step verification and save → Vertical calibration → Vertical verification and save → Final print check.
Do not skip step calibration and do vertical calibration directly.
Note: Poor nozzle status can make the calibration result inaccurate.
For the nozzle test pass standard and guide location, see [Test Strip Printing (Nozzle Test) Guide].
Double-click the PM icon on the computer desktop to start the Print Manager software (See Fig. 3.1).
Click Calibrate on the menu bar (See Fig. 3.2).
Important: Save a screenshot of the original parameters before changing any value.
If the calibration result becomes worse, restore the original settings.
Check Criteria:
| Result | Action |
|---|---|
| Two lines fully overlap with no gap | Select this group |
| Lines are offset, have shadows, or do not touch | Do not select this group |
| Several groups are close to overlapping | Select the thinnest and clearest group |
①:First Row
②:One Group of Lines
Example: If the fully overlapping group value is -3, enter -3 in Revise, then click => (See Fig. 4.3).
Direction Check:
| Result | Action |
|---|---|
| Lines are closer to overlapping | The adjustment direction is correct |
| Lines are farther from overlapping | Adjust in the opposite direction |
Important: If you do not click Save, the step calibration result will not take effect.
Important: Start vertical calibration only after step calibration is complete and Save has been clicked.
Reading Rule:
| What You See | What to Do |
|---|---|
| The best overlap is at 0 | No adjustment is needed for this row |
| The best overlap is not at 0 | Record that value for this row |
| Several groups look close | Select the thinnest and clearest group |
Example:
If the best overlap in row G1(C) is under -3, record:
| Field Name | Recorded Value |
|---|---|
| G1(C) | -3 |
Use this formula:
Final Value = Original Value + Recorded Value
Example: G1(C) means row G1 and column (C).
This number is the Original Value.
Calculation Examples:
| Field Name | Original Value | Recorded Value | Final Value |
|---|---|---|---|
| Example G1(C) | 3 | -3 | 0 |
| Example G1(Y) | -5 | -2 | -7 |
| Example 3 | 2 | 3 | 5 |
| Example 4 | -2 | 3 | 1 |
Example for G1(C):
If the original value of G1(C) is 3, and the recorded value from the pattern is -3:
3 + (-3) = 0
Enter 0 in the G1(C) field.
Important: Keep the plus or minus sign.
Check the field name and final value before entry.
Note: If multiple rows need adjustment, adjust them row by row.
Follow this sequence:
Adjust one row → Print and verify → Confirm that this row is complete → Adjust the next row.
Important: If do not click Save, the vertical calibration result will not take effect.
| Check Item | Pass Criteria |
|---|---|
| Vacuum suction | Vacuum suction is on. Material does not slip |
| Material | Flat. Parallel to paper feeding direction. No skewing or slipping |
| Feeder | Forward / advance state |
| Nozzle test | No obvious broken lines, missing nozzles, or missing colors |
If the issue is still not solved after checking again and recalibrating, contact LINKO Technical Support.
Warning: Do not continue to change parameters randomly.
LINKO Technical Support
Email: support@dtflinko.com
Website: www.dtflinko.com
© 2026 LINKO Technical Team. All rights reserved.
Reproduction or modification of this document without permission is prohibited.
Applicable Model: K-602, K-603, K-604KM-602, S-30, S-60A-602, A-604, A-502
(More models may be added in future updates.)
Goal: Fix vertical misalignment between the white and color layers so they are perfectly aligned top-to-bottom.
❌ Incorrect
White and color layers not aligned — please adjust
✅ Correct
White and color layers fully aligned — no need to adjust
1.1 Check and Turn ON Vacuum Suction
1.2 Check the Paper Feeder Switch — set to Forward
1.3 Check the Feed motor damping system — Left and right sides are in relaxed state.
1.4 Perform Two Calibrations:
1.1 Check and Turn ON Vacuum Suction
1.2 Perform Two Calibrations:
Solution and Step-by-Step Instructions
1.1-1-Location-of-Suction-Switch
1.1-1-Location-of-Suction-Switch
❌ Incorrect
1.1-2 Suction closed
✅ Correct
1.1-3 Suction on
1.2-1 Paper feeder switch
❌ Incorrect
1.2-2-Reverse
1.2-3-Of
✅Correct
1.2-4-Forward
Step-by-Step Instructions:
Step 1 Access the Calibration Menu
Open the printer control software and click the “Calibrate” section.
Enter the calibration page
Step 2 Access and Print the Calibration Chart
A dialog box will appear. Select “Base Step”, then click “Print” to start printing the step calibration chart. (Ensure the media is loaded and the printer is ready)
1.2.1-2-Print-the-Step-Calibration-Chart
Step 3 Find the set of lines where the left and right columns overlap most clearly, and Record the number.
Look at the printed chart. In the first row, find the group of lines from the left and right columns that overlap perfectly, and record the number shown below.
1.2.1-3-Find-the-Lines-that-Overlaps-Perfectly
Step 4 Adjust the Calibration value
Enter the number in “Revise” and click “=>”
Use the value that gives the best overlap — for example, if -32 looks best, enter -32 as the Calibration number.
If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.91.1, 1.2… up to 1.9, until the best result is achieved.
1.2.1-4-Adjust-the-Calibration-Value
Step 5 Save the Calibration value
Click “Save” to apply it.
1.2.1-5-Save-the-Calibration-Value
Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.
⚠️ If the value is still not “0” after several attempts, stop and check:
Fix these issues before repeating calibration.
1.2.1-6-Step-Calibration-Completed
Ensures white and color layers align correctly in the vertical direction.
Step-by-Step Instructions:
Step 1: Access the Vertical Calibration Menu
Open the printer control software and click the “Calibrate” tab.
1.2.2-1
Step 2 Access and Print the Vertical Calibration Chart
A dialog box pops up. Click “Vertical/Overlap”→Click “VertCal Calibration”→Click “Print” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready)
1.2.2-2-Print-the-Vertical-Calibration-Chart
Step 3 Find the group of lines
That Overlaps Perfectly between the left and right columns, and record the number below.
1.2.2-3-Find-the-Lines-that-Overlaps-Perfectly
Step 4 Calculate and Set the Final Calibration Value
Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 35 and the overlap value is -3-5, then final value:
3-3 + (-3) =05 + (-5) =0
Enter 0 into the G1(C)G2, 1(W1, W2, W3, W4, W5, W6, W7, W8) parameter field in the system. (No need to modify the number in other parameter fields)
1.2.2-4-Set-the-Calibration-Value
Step 5 Save the Calibration Settings
Click “Save” to apply the new calibration value.
1.2.2-5-Save the Calibration Settings
Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.
⚠️If the value is still not “0” after several attempts, stop and check:
Correct these issues before redoing calibration.
1.2.2-6-Vertical-Calibration-Completed
1.3-1 Feed motor damping system
❌ Incorrect
1.3-2-Feed-motor-damping-relaxed
✅ Correct
1.3-3-Feed-motor-damping-tightened
✅ Relaxed State: The screw is almost fully tightened inward, with little to no thread visible.
❌ Tightened State: Multiple threads are visibly exposed, indicating the screw is not fully tightened.
1.4.1 Step Calibration: Adjusts the feed distance to ensure accurate image scaling.
Step-by-Step Instructions:
Step 1 Access the Calibration Menu
Open the printer control software and click the “Calibrate” tab.
1.4.1-1-Access-the-Step-Calibration-Menu
Step 2 Access and Print the Calibration Chart
A dialog box pops up. Click “Base Step”, then click “Print” to start printing the step calibration chart.(Ensure the media is loaded and the printer is ready.)
1.4.1-2-Print-the-Step-Calibration-Chart
Step 3 Find the group of lines that overlaps perfectly between the left and right columns, and record the number.
Look at the printed chart. In the first row, find the group of lines from the left and right columns that overlap perfectly, and record the number.
1.4.1-3-Find-the-Lines-that-Overlaps-Perfectly
Step 4 Adjust the Calibration value
Enter the number in “Revise” and click “=>”
Use the value that gives the best overlap — for example, if -3 looks best, enter -3 as the Calibration number.
If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.9, until the best result is achieved.
1.4.1-4-Adjust-the-Calibration-Value
Step 5 Save the Calibration value
Click “Save” to apply it.
1.4.1-5-Save-the-Calibration-Value-K602-K603-K604
Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.
⚠ If the value is still not “0” after several attempts, stop and check:
Fix these issues before repeating calibration.
1.4.1-6-Step-Calibration-Completed
1.4.2 Vertical Alignment Calibration: Ensures white and color layers align correctly in the vertical direction.
Step-by-Step Instructions:
Step 1 Access the Vertical Calibration Menu
Open the printer control software and click the “Calibrate” tab.
1.4.2-1-Access-the-Vertical-Calibration-Menu
Step 2 Access and Print the Vertical Calibration Chart
A dialog box pops up. Click “Vertical/Overlap”→Click “VertCal Calibration”→Click “Print ” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready.)
1.4.2-2-Print-the-Vertical-Calibration-Chart
Step 3 Find the group of lines That Overlaps Perfectly between the left and right columns, and record the number below.
1.4.2-3-Record-the-Number
Step 4 Calculate and Set the Final Calibration Value
Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 5 and the overlap value is -5, then the final value:
5 + (-5) =0
Enter 0 into the G2, 1(W1,W2,W3,W4,W5,W6,W7,W8) parameter field in the system. (No need to modify the number in other parameter fields)
1.4.2-4-Set-the-Calibration-Value
Step 5 Save the Calibration Settings
Click “Save” to apply the new calibration value.
1.4.2-5-Save the Calibration Settings
Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.
⚠ If the value is still not “0” after several attempts, stop and check:
Correct these issues before redoing calibration.
1.4.2-6-Vertical-Calibration-Completed
The white and color layers are not properly aligned in the vertical direction.
Incorrect❌
White and color layers not aligned — please adjust
Correct✅
White and color layers fully aligned — no need to adjust
(This involves two steps: step calibration first, followed by vertical alignment calibration).
4.1 Step Calibration: Adjusts the feed distance to ensure accurate image scaling.
Step-by-Step Instructions:
4.1.1 Access the Calibration Menu
Open the printer control software and Click the “Adjust” section.
4.1.2 Access and Print the Calibration Chart
Click “Step Adjust”, then click “Print Adjust” to start printing the step calibration chart.
4.1.3 Find the Line That Overlaps Perfectly and Record the number.
Look at the printed chart and find the line where the two sets overlap most clearly. Write down the number shown next to that line.
4.1.4 Adjust the Calibration value
Enter the number and click “Calculate”
Step ①: Rough Adjustment (Initial Calibration)
Use the value that gives the best overlap — for example, if -3 looks best, enter -3 as the calibration number.
Step ②: Fine Adjustment (Precise Calibration)
If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.9, until the best result is achieved.
4.1.5 Enter and Save the Calibration value
Input the final calibration number and click “Save” to apply it.
4.1.6 Check if Calibration is Complete
If the number where both left and right lines overlap is “0”, calibration is complete. If not, repeat the above steps until it is.
4.2 Vertical Alignment Calibration: Ensures white and color layers align correctly in the vertical direction.
Step-by-Step Instructions:
Step-by-Step Instructions:
4.2.1 Access the Vertical Calibration Menu
Open the printer control software and Click “Adjust”section.
4.2.2 Access the Vertical Calibration Page
Click “Head Space”→Click “Vertical”→Click “Print Adjust” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready)
4.2.3 Find the Value Where Black and White Overlap
Look at the printed chart and identify the number where the black line and white background overlap most perfectly. Record that number.
4.2.4 Calculate and Set the Final Calibration Value
Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 731.8 and the overlap value is -4, then:
731.8 + (-4) = 727.8
Enter 727.8 into the H2 parameter field in the system. (No need to modify the number in H1)
4.2.5 Save the Calibration Settings
Click “Save” to apply the new calibration value.
If the overlap value is 0, the calibration is complete. If not, repeat the steps until it is.