Select Your Printer Model
(Default: k-605 DTF Printer)

Choose your model from the drop-down menu below.↓↓↓


LINKO K605 Step and Vertical Calibration Guide

Applicable model: LINKO K605 DTF Printer

This guide is used to solve vertical misalignment or ghosting between the white ink layer and the color ink layer.

1. Check Whether Calibration Is Needed First

Calibration is needed when the white ink layer and the color ink layer are not aligned vertically in the printed image.

Condition Check Result Next Step
White ink layer and color ink layer are fully aligned Normal No calibration needed
Vertical misalignment or ghosting between white ink layer and color ink layer Abnormal Do step calibration first. Then do vertical calibration
Broken nozzle lines, missing nozzles, obvious missing colors, or material slipping Direct calibration is not recommended Check nozzle status, material, and film feeding first
Fig.1-1 White and Color Layer Alignment Check

1-1 White and Color Layer Alignment Check

Fig.1-2 Vertical Misalignment Check

1-2 Vertical Misalignment Check

Important: Follow this sequence:

Pre-check → Step calibration → Step verification and save → Vertical calibration → Vertical verification and save → Final print check.

Do not skip step calibration and do vertical calibration directly.

2. Preparation Before Start

2.1 Turn On Vacuum Suction

  • Turn on the vacuum suction function (See Fig. 2.1-1).
  • Confirm that the material can be firmly held on the platform.
  • If the material curls up, slides, or is not held evenly, solve the material or vacuum suction issue first.
Fig.2.1-1 Vacuum Suction Location

2.1-1 Vacuum Suction Location

Fig.2.1-2 Vacuum Suction Status

2.1-2 Vacuum Suction Status

2.2 Confirm the Feeding Direction

  • Set the feeder switch to Forward (See Fig. 2.2-1).
  • Confirm that the material feeds forward.
  • If the feeder is in reverse state or off state, set it to forward first.
Fig.2.2-1 Feeder switch Location

2.2-1-Feeder switch Location

Fig.2.2-2 Confirm the Feeding Direction

2.2-2 Confirm the Feeding Direction

2.3 Check the Material

  • Confirm that the material surface is flat.
  • Confirm that the material edge is parallel to the paper feeding direction.
  • Confirm that the material has no wrinkles, curled edges, slipping, or skewing (See Fig. 2.3).
Fig.2.3 Flat Material Placement

2.3 Flat Material Placement

2.4 Check the Nozzle Status

  • Print the nozzle test pattern.
  • If there are continuous broken lines or missing nozzles, clean the nozzle first.
  • Start calibration after the nozzle status returns to normal.

Note: Poor nozzle status can make the calibration result inaccurate.

For the nozzle test pass standard and guide location, see [Test Strip Printing (Nozzle Test) Guide].

3. Open the Calibration Screen and Back Up the Original Parameters

3.1 Start the Print Manager Software

Double-click the PM icon on the computer desktop to start the Print Manager software (See Fig. 3.1).

Fig.3.1 Double Click PM Icon

3.1 Double Click PM Icon

3.2 Open Calibrate

Click Calibrate on the menu bar (See Fig. 3.2).

Fig.3.2 Click Calibrate

3.2 Click Calibrate

3.3 Back Up the Original Parameters

  • After entering the Calibrate window, click Base Step.
  • Take a screenshot to save the current step calibration parameters.
  • Switch to the Vertical/Overlap page.
  • Take a screenshot to save the current vertical calibration parameters (See Fig. 3.3).

Important: Save a screenshot of the original parameters before changing any value.

If the calibration result becomes worse, restore the original settings.

Fig.3.3 Original Parameter Backup

3.3 Original Parameter Backup

4. Step Calibration

4.1 Print the Step Calibration Pattern

  • Confirm that the material is placed flat.
  • Confirm that vacuum suction is turned on.
  • Confirm that the feeder is in forward state.
  • In the Calibrate window, click Base Step.
  • Click Print(See Fig. 4.1).
  • Wait until the step calibration pattern is printed.
Fig.4.1 Click Base Step Print

4.1 Click Base Step Print

4.2 Read the Step Calibration Value

  • Check only the top row of calibration lines on the step calibration pattern.
  • Compare each group of lines from left to right.
  • Find the group where the two lines fully overlap with no gap.
  • Record the value below that group (See Fig. 4.2).

Check Criteria:

Result Action
Two lines fully overlap with no gap Select this group
Lines are offset, have shadows, or do not touch Do not select this group
Several groups are close to overlapping Select the thinnest and clearest group
Fig.4.2-1 How to Locate the Calibration Lines

4.2-1 How to Locate the Calibration Lines

①:First Row
②:One Group of Lines

Fig.4.2 2 Step Calibration Result Check

4.2-2 Step Calibration Result Check

4.3 Enter the Step Calibration Value

  • Enter the recorded value in the Revise input box.
  • Click =>.

Example: If the fully overlapping group value is -3, enter -3 in Revise, then click => (See Fig. 4.3).

Fig.4.3 Step Calibration Value Entry

4.3 Step Calibration Value Entry

4.4 Reprint and Verify the Step Result

  • Reprint the step calibration pattern.
  • Check whether the left and right lines at the 0 position fully overlap.
  • If the 0 position fully overlaps, step calibration is complete (See Fig. 4.4-1).
  • If the 0 position does not fully overlap, fine-tune by 0.1 based on the current value (See Fig. 4.4-2).
  • After each fine-tune, repeat Enter → Print → Check.

Direction Check:

Result Action
Lines are closer to overlapping The adjustment direction is correct
Lines are farther from overlapping Adjust in the opposite direction
Fig.4.4-1 Zero Position Overlap Check

4.4-1 Zero Position Overlap Check

Fig.4.4-2 Zero Position Misalignment Check

4.4-2 Zero Position Misalignment Check

4.5 Save the Step Parameters

  • Confirm that the 0 position fully overlaps.
  • Click Save(See Fig. 4.5).

Important: If you do not click Save, the step calibration result will not take effect.

Fig.4.5 Click Save

4.5 Click Save

5. Vertical Calibration

Important: Start vertical calibration only after step calibration is complete and Save has been clicked.

5.1 Print the Vertical Calibration Pattern

  • Stay in the Calibrate window.
  • Click Vertical/Overlap.
  • Select VertCal Calibration.
  • Click Print.
  • Wait until the vertical calibration pattern is printed (See Fig. 5.1).
Fig.5.1 Click Vertical Overlap VertCal Calibration Print

5.1 Click Vertical Overlap VertCal Calibration Print

5.2 Read the Vertical Calibration Values Row by Row

  • Start from the first row.
  • Check one row at a time.
  • Find the group where the left and right lines overlap best.
  • Record the value below that group.
  • Repeat the same check for each row that needs adjustment (See Fig. 5.2).

Reading Rule:

What You See What to Do
The best overlap is at 0 No adjustment is needed for this row
The best overlap is not at 0 Record that value for this row
Several groups look close Select the thinnest and clearest group

Example:

If the best overlap in row G1(C) is under -3, record:

Field Name Recorded Value
G1(C) -3
Fig.5.2 Read the Recorded Value from One Row

5.2 Read the Recorded Value from One Row

5.3 Calculate and Enter the Vertical Calibration Values

Use this formula:

Final Value = Original Value + Recorded Value

  • Find the field name in the software.

  Example: G1(C) means row G1 and column (C).

  • Check the original value shown in that field.

This number is the Original Value.

  • Add the recorded value from the calibration pattern.
  • Enter the final value in the same field (See Fig. 5.3).
  • Change only the fields that need adjustment.

Calculation Examples:

Field Name Original Value Recorded Value Final Value
Example G1(C) 3 -3 0
Example G1(Y) -5 -2 -7
Example 3 2 3 5
Example 4 -2 3 1

Example for G1(C):

If the original value of G1(C) is 3, and the recorded value from the pattern is -3:

3 + (-3) = 0

Enter 0 in the G1(C) field.

Important: Keep the plus or minus sign.

Check the field name and final value before entry.

Fig.5.3-1 Locate the Field and Enter the Final Value(Example G1(C))

5.3-1 Locate the Field and Enter the Final Value(Example G1(C))

Fig.5.3-2 Locate the Field and Enter the Final Value(Example G1(Y))

5.3-2 Locate the Field and Enter the Final Value(Example G1(Y))

5.4 Reprint and Verify Vertical Calibration

  • Reprint the vertical calibration pattern.
  • Check whether the 0 position of the adjusted row fully overlaps.
  • If the 0 position fully overlaps, calibration for that row is complete.
  • If it does not fully overlap, fine-tune by ±0.1 based on the current value.
  • After each fine-tune, reprint and check again.

Note: If multiple rows need adjustment, adjust them row by row.

Follow this sequence:

Adjust one row → Print and verify → Confirm that this row is complete → Adjust the next row.

Fig.5.4-1 Row Zero Position Overlap Check

5.4-1 Row Zero Position Overlap Check

Fig.5.4-2 Row Zero Position Misalignment Check

5.4-2 Row Zero Position Misalignment Check

5.5 Save the Vertical Calibration Parameters

  • Confirm that the 0 position fully overlaps for all rows that need adjustment.
  • Click Save in the lower right corner.

Important: If do not click Save, the vertical calibration result will not take effect.

Fig.5.5 Click Save

5.5 Click Save

6. Final Check

  • Reprint the pattern that needs to be checked.
  • Confirm whether the white ink layer and the color ink layer are fully aligned.
  • If they are fully aligned, calibration is complete.
  • If vertical misalignment or ghosting still exists, check again according to the table below.
Check Item Pass Criteria
Vacuum suction Vacuum suction is on. Material does not slip
Material Flat. Parallel to paper feeding direction. No skewing or slipping
Feeder Forward / advance state
Nozzle test No obvious broken lines, missing nozzles, or missing colors

If the issue is still not solved after checking again and recalibrating, contact LINKO Technical Support.

Warning: Do not continue to change parameters randomly.

LINKO Technical Support
Email: support@dtflinko.com
Website: www.dtflinko.com

© 2026 LINKO Technical Team. All rights reserved.
Reproduction or modification of this document without permission is prohibited.

Troubleshooting Print Alignment and Ghosting Issues – Version DVersion EVersion A

Applicable Model: K-602, K-603, K-604KM-602, S-30, S-60A-602, A-604, A-502

(More models may be added in future updates.)

White and Color Layers Not Aligned Vertically — How to Calibrate? (Vertical Misalignment – White/Color Ink)

Goal: Fix vertical misalignment between the white and color layers so they are perfectly aligned top-to-bottom.

❌ Incorrect

Vertical Misalignment Example

White and color layers not aligned — please adjust

✅ Correct

Vertical Alignment Example

White and color layers fully aligned — no need to adjust

Full Process (4 Main Tasks):

1.1 Check and Turn ON Vacuum Suction

1.2 Check the Paper Feeder Switch — set to Forward

1.3 Check the Feed motor damping system — Left and right sides are in relaxed state.

1.4 Perform Two Calibrations:

  • 1.4.1 Step Calibration (adjusts feed distance for accurate scaling)
  • 1.4.2 Vertical Alignment Calibration (aligns white and color layers vertically)

Full Process (2 Main Tasks):

1.1 Check and Turn ON Vacuum Suction

1.2 Perform Two Calibrations:

  • 1.2.1 Step Calibration (adjusts feed distance for accurate scaling)
  • 1.2.2 Vertical Alignment Calibration (aligns white and color layers vertically)

Detailed Steps

Solution and Step-by-Step Instructions

1.1 Make sure the vacuum suction is turned on.

Location of Suction Switch K602 K 603 K 604

1.1-1-Location-of-Suction-Switch

Location of Suction Switch KM602 S30 S60

1.1-1-Location-of-Suction-Switch

❌ Incorrect

Labeling of Suction closed

1.1-2 Suction closed

✅ Correct

Labeling of Suction on

1.1-3 Suction on

1.2 Paper feeder switch — set to ‘Forward’

Location of Paper Feeder Switch K602 K603 K604

1.2-1 Paper feeder switch

❌ Incorrect

Labeling of Paper Feeder Switch Reverse K602 K603 K604

1.2-2-Reverse

Labeling of Paper Feeder Switch Off K602 K603 K604

1.2-3-Of

✅Correct

Labeling of Paper Feeder Switch Forward K602 K603 K604

1.2-4-Forward

1.31.2 Check the calibration settings (this involves two steps: step calibration first, followed by vertical alignment calibration).

1.3.11.2.1 Step Calibration: Adjusts the feed distance to ensure accurate image scaling.

Step-by-Step Instructions:

Step 1 Access the Calibration Menu
Open the printer control software and click the “Calibrate” section.

Access the Step Calibration Menu

Enter the calibration page

Step 2 Access and Print the Calibration Chart
A dialog box will appear. Select “Base Step”, then click “Print” to start printing the step calibration chart. (Ensure the media is loaded and the printer is ready)

Print the Step Calibration Chart KM602 S30 S60

1.2.1-2-Print-the-Step-Calibration-Chart

Step 3 Find the set of lines where the left and right columns overlap most clearly, and Record the number.

Look at the printed chart. In the first row, find the group of lines from the left and right columns that overlap perfectly, and record the number shown below.

Find the Lines that Overlaps Perfectly KM602 S30 S60

1.2.1-3-Find-the-Lines-that-Overlaps-Perfectly

Step 4 Adjust the Calibration value

Enter the number in “Revise” and click “=>”

  • Rough Adjustment (Initial Calibration)

Use the value that gives the best overlap — for example, if -32 looks best, enter -32 as the Calibration number.

  • Fine Adjustment (Precise Calibration)

If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.91.1, 1.2… up to 1.9, until the best result is achieved.

Adjust the Calibration Value KM602 S30 S60

1.2.1-4-Adjust-the-Calibration-Value

Step 5 Save the Calibration value

Click “Save” to apply it.

Save the Calibration Value KM602 S30 S60

1.2.1-5-Save-the-Calibration-Value

Step 6 Check if Calibration is Complete

If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.

⚠️ If the value is still not “0” after several attempts, stop and check:

  • Is vacuum suction turned on?
  • Is the DTF film slipping or not properly loaded?

Fix these issues before repeating calibration.

Step Calibration Completed KM602 S30 S60

1.2.1-6-Step-Calibration-Completed

1.3.21.2.2 Vertical Alignment Calibration

Ensures white and color layers align correctly in the vertical direction.

Step-by-Step Instructions:

Step 1: Access the Vertical Calibration Menu
Open the printer control software and click the “Calibrate” tab.

Access the Vertical Calibration Menu

1.2.2-1

Step 2 Access and Print the Vertical Calibration Chart
A dialog box pops up. Click “Vertical/Overlap”→Click “VertCal Calibration”→Click “Print” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready)

Print the Vertical Calibration Chart KM602 S30 S60

1.2.2-2-Print-the-Vertical-Calibration-Chart

Step 3 Find the group of lines

That Overlaps Perfectly between the left and right columns, and record the number below.

  • First, look at the printed chart. Each row of calibration lines has its name indicated on the left side of the line. For example, the calibration line in the first row is labeled G1 (K)G1,1(C, M, Y, K, G, R, B, O), the second row is labeled G1 (C)G2, 1(W1, W2, W3, W4, W5, W6, W7, W8), and so on.
  • Then, for each row, find the group of lines from the left and right columns that are perfectly overlapping, and check the number displayed directly below that group.
    If the value is not “0”, you need to adjust that row accordingly.
  • For example, in the second row, the value below the group of lines that overlap perfectly on the left and right columns is -3″-5, indicating that this row needs adjustment.
  • If all other rows have an overlapping group with a value of “0”, then those rows do not need adjustment.
  • Finally, record the name and value of the row that requires adjustment:G1 (C)G2, 1 (W1, W2, W3, W4, W5, W6, W7, W8) and -3-5.
Find the Lines that Overlaps Perfectly KM602 S30 S60

1.2.2-3-Find-the-Lines-that-Overlaps-Perfectly

Step 4 Calculate and Set the Final Calibration Value

Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 35 and the overlap value is -3-5, then final value:
3-3 + (-3) =05 + (-5) =0
Enter 0 into the G1(C)G2, 1(W1, W2, W3, W4, W5, W6, W7, W8) parameter field in the system. (No need to modify the number in other parameter fields)

Set the Calibration Value KM602 S30 S60

1.2.2-4-Set-the-Calibration-Value

Step 5 Save the Calibration Settings
Click “Save” to apply the new calibration value.

Save the Calibration Settings KM602 S30 S60

1.2.2-5-Save the Calibration Settings

Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.

⚠️If the value is still not “0” after several attempts, stop and check:

  • Is vacuum suction turned on?
  • Is the DTF film flat and parallel to the feed path?
  • Is the DTF film slipping or not properly loaded?

Correct these issues before redoing calibration.

Vertical Calibration Completed KM602 S30 S60

1.2.2-6-Vertical-Calibration-Completed

1.3 Feed motor damping system — Left and right sides are in relaxed state.

Feed motor damping system K602 K603 K604

1.3-1 Feed motor damping system

❌ Incorrect

Feed motor damping relaxed K602 K603 K604

1.3-2-Feed-motor-damping-relaxed

✅ Correct

Feed motor damping tightened K602 K603 K604

1.3-3-Feed-motor-damping-tightened

✅ Relaxed State: The screw is almost fully tightened inward, with little to no thread visible.
❌ Tightened State: Multiple threads are visibly exposed, indicating the screw is not fully tightened.

1.4 Check the calibration settings (this involves two steps: step calibration first, followed by vertical alignment calibration).

1.4.1 Step Calibration: Adjusts the feed distance to ensure accurate image scaling.

Step-by-Step Instructions:

Step 1 Access the Calibration Menu
Open the printer control software and click the “Calibrate” tab.

Access the Step Calibration Menu

1.4.1-1-Access-the-Step-Calibration-Menu

Step 2 Access and Print the Calibration Chart
A dialog box pops up. Click “Base Step”, then click “Print” to start printing the step calibration chart.(Ensure the media is loaded and the printer is ready.)

Print the Step Calibration Chart

1.4.1-2-Print-the-Step-Calibration-Chart

Step 3 Find the group of lines that overlaps perfectly between the left and right columns, and record the number.
Look at the printed chart. In the first row, find the group of lines from the left and right columns that overlap perfectly, and record the number.

Find the Lines that Overlaps Perfectly

1.4.1-3-Find-the-Lines-that-Overlaps-Perfectly

Step 4 Adjust the Calibration value
Enter the number in “Revise” and click “=>”

  • Rough Adjustment (Initial Calibration)

Use the value that gives the best overlap — for example, if -3 looks best, enter -3 as the Calibration number.

  • Fine Adjustment (Precise Calibration)

If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.9, until the best result is achieved.

Adjust the Calibration Value

1.4.1-4-Adjust-the-Calibration-Value

Step 5 Save the Calibration value
Click “Save” to apply it.

Save the Calibration Value

1.4.1-5-Save-the-Calibration-Value-K602-K603-K604

Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.

⚠ If the value is still not “0” after several attempts, stop and check:

  • Is vacuum suction turned on?
  • Is the paper feeder set to Forward?
  • Is the feed motor damping system on left and right sides are in relaxed state?
  • Is the DTF film slipping or not properly loaded?

Fix these issues before repeating calibration.

Step Calibration Completed

1.4.1-6-Step-Calibration-Completed

1.4.2 Vertical Alignment Calibration: Ensures white and color layers align correctly in the vertical direction.

Step-by-Step Instructions:

Step 1 Access the Vertical Calibration Menu
Open the printer control software and click the “Calibrate” tab.

Access the Step Calibration Menu

1.4.2-1-Access-the-Vertical-Calibration-Menu

Step 2 Access and Print the Vertical Calibration Chart
A dialog box pops up. Click “Vertical/Overlap”→Click “VertCal Calibration”→Click “Print ” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready.

Print the Vertical Calibration Chart K602 K603 K604

1.4.2-2-Print-the-Vertical-Calibration-Chart

Step 3 Find the group of lines That Overlaps Perfectly between the left and right columns, and record the number below.

  • First, look at the printed chart. Each row of calibration lines has its name indicated on the left side of the line. For example, the calibration line in the first row is labeled “G1, 1 (C, M, Y, K, G, R, O, B)”, the second row is labeled “G2,1(W1, W2, W3, W4, W5, W6, W7, W8)”, and so on.
  • Then, for each row, find the group of lines from the left and right columns that are perfectly overlapping, and check the number displayed directly below that group.
    If the value is not “0”, you need to adjust that row accordingly.
  • For example, in the second row, the value below the group of lines that overlap perfectly on the left and right columns is “-5“, indicating that this row needs adjustment.
  • If all other rows have an overlapping group with a value of “0”, then those rows do not need adjustment.
  • Finally, record the name and value of the row that requires adjustment: “G2, 1 (W1, W2, W3, W4, W5, W6, W7, W8)” and “-5”.
Record the Number K602 K603 K604

1.4.2-3-Record-the-Number

Step 4 Calculate and Set the Final Calibration Value
Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 5 and the overlap value is -5, then the final value:
5 + (-5) =0
Enter 0 into the G2, 1(W1,W2,W3,W4,W5,W6,W7,W8) parameter field in the system. (No need to modify the number in other parameter fields)

Set the Calibration Value K602 K603 K604

1.4.2-4-Set-the-Calibration-Value

Step 5 Save the Calibration Settings
Click “Save” to apply the new calibration value.

Save the Calibration Settings K602 K603 K604

1.4.2-5-Save the Calibration Settings

Step 6 Check if Calibration is Complete
If the value of the overlapping line group is “0”, calibration is complete. If not, repeat the above steps until it is.
⚠ If the value is still not “0” after several attempts, stop and check:

  • Is the suction working properly?
  • Is the DTF film flat and parallel to the feed path?
  • Is the paper feeder switch in Forward?
  • Is the DTF film slipping or skewed?

Correct these issues before redoing calibration.

Vertical Calibration Completed K602 K603 K604

1.4.2-6-Vertical-Calibration-Completed

Vertical Misalignment Between White and Color Inks

The white and color layers are not properly aligned in the vertical direction.

Vertical misalignment Error example

Incorrect❌

White and color layers not aligned — please adjust

Vertical misalignment Correct example

Correct✅

White and color layers fully aligned — no need to adjust

Solution and Step-by-Step Instructions:

Step 1: Make sure the vacuum suction is turned on.

Labeling of suction systems

Step 2: Paper feeder switch — set to ‘Forward’

Feed switch labeling

Step 3: Check the tension knobs on both sides of the paper feeding motors. Make sure both are in a relaxed state.

  • Relaxed State: The knob is almost fully screwed in, with little or no thread exposed.
  • Tightened State: The knob is less screwed in, with several thread loops clearly visible.
Damping labeling

Step 4: Check the calibration settings

(This involves two steps: step calibration first, followed by vertical alignment calibration).

4.1 Step Calibration: Adjusts the feed distance to ensure accurate image scaling.

Step-by-Step Instructions:

4.1.1 Access the Calibration Menu
Open the printer control software and Click the “Adjust” section.

Access the Calibration Menu

4.1.2 Access and Print the Calibration Chart
Click “Step Adjust”, then click “Print Adjust” to start printing the step calibration chart.

Access and Print the Calibration Chart

4.1.3 Find the Line That Overlaps Perfectly and Record the number.
Look at the printed chart and find the line where the two sets overlap most clearly. Write down the number shown next to that line.

Find the Line That Overlaps Perfectly

4.1.4 Adjust the Calibration value
Enter the number and click “Calculate”

Step ①: Rough Adjustment (Initial Calibration)
Use the value that gives the best overlap — for example, if -3 looks best, enter -3 as the calibration number.

Step ②: Fine Adjustment (Precise Calibration)
If the result is still not ideal, make fine adjustments using decimal number, such as -2.1, -2.2… up to -2.9, until the best result is achieved.

Adjust the Calibration value

4.1.5 Enter and Save the Calibration value
Input the final calibration number and click “Save” to apply it.

Enter and Save the Calibration value

4.1.6 Check if Calibration is Complete
If the number where both left and right lines overlap is “0”, calibration is complete. If not, repeat the above steps until it is.

Calibration complete

4.2 Vertical Alignment Calibration: Ensures white and color layers align correctly in the vertical direction.
Step-by-Step Instructions:

Step-by-Step Instructions:

4.2.1 Access the Vertical Calibration Menu
Open the printer control software and Click “Adjust”section.

Access the Calibration Menu

4.2.2 Access the Vertical Calibration Page
Click “Head Space”→Click “Vertical”→Click “Print Adjust” to print the vertical calibration chart (Ensure the media is loaded and the printer is ready)

Access the Vertical Calibration Page

4.2.3 Find the Value Where Black and White Overlap
Look at the printed chart and identify the number where the black line and white background overlap most perfectly. Record that number.

Find the Value Where Black and White Overlap

4.2.4 Calculate and Set the Final Calibration Value
Use the following formula:
Final Value = Original Value + Overlap Value
Example: If the original value is 731.8 and the overlap value is -4, then:
731.8 + (-4) = 727.8
Enter 727.8 into the H2 parameter field in the system. (No need to modify the number in H1)

  • Rough Adjustment: Start with the integer value where overlap is most obvious.
  • Fine Adjustment: If needed, fine-tune using decimal values (e.g. -0.2, +0.3, etc.) until the overlap is perfect.
Calculate and Set the Final Calibration Value

4.2.5 Save the Calibration Settings
Click “Save” to apply the new calibration value.

Save the Calibration Settings

If the overlap value is 0, the calibration is complete. If not, repeat the steps until it is.

Vertical calibration complete